Your Location:Home > PRODUCTS>Protective Material Testing Equipments>Mask Testing Equipments

XHF-107 Bacterial Filtration Efficiency (BFE) Tester

Product Description of XHF-107 Bacterial Filtration Efficiency (BFE) Tester

Medical Face Masks Bacterial Filtration Efficiency (BFE) Tester is suitable for metrological verification departments, scientific research institutes, mask manufacturers and other relevant departments to measure the bacterial filtration efficiency (BFE) of medical face mask materials. On the basis of complying related standards, its innovative dual air paths for simultaneous contrast sampling enhance the accuracy of sampling.


Test Standards of XHF-107 Bacterial Filtration Efficiency (BFE) Tester

YY/T 0969-2013, YY 0469-2011, ASTM F 2100, ASTM F 2101, EN 14683, CNS 14775


Product Features of XHF-107 Bacterial Filtration Efficiency (BFE) Tester

Negative pressure test system to ensure operator safety.

 

Negative pressure cabinet built-in peristaltic pump, A,B two path 6 Grade Andersen sampling.

 

Peristaltic pump flow is adjustable. 

 

Special microbial aerosol generator, the spray volume of bacterial liquid can be set, and the atomization effect is good.

 

Built-in high speed industry micro computer control.

 

Adopt 10 Inch industry high resolution color touch screen.

 

USB Interface for data stroage.

 

Cabiney built in high brightness lighting.

 

Built in electricity leakage protective switch to ensure operator safety.

 

The inner layer of cabinet is made of formed stainless steel,the outer layer is sprayed with cold-rolled plates, between the inner and outer layers are insulated, flame retardant.

.


Technical Parameter of XHF-107 Bacterial Filtration Efficiency (BFE) Tester

Item

Parameter Range

Resolution

Max Permissible Errors

Sampling flow rate of Path A

28.3L/min

0.01L/min

2.5%

Sampling flow rate of Path B

28.3L/min

0.01L/min

2.5%

Nebulization flow rate

(8~10)L/min

0.1L/min

5.0%

Peristaltic pump flow rate

0.0063mL/min

0.001mL/min

2.5%

Flow meter fore pressure of path A

-200kPa

0.01kPa

2.5%

Flow meter fore pressure of path B

-200kPa

0.01kPa

2.5%

Nebulization flow meter fore pressure

0300kPa

0.1kPa

2.5%

Negative pressure of aerosol chamber

-5000Pa

0.1Pa

2.0%

Negative pressure of cabinet body

(-500-0)Pa

Data Storage

>100000 Groups

Feature of HEPA filter

The filtration efficiency of particles above 0.3um is ≥99.99%

Mass mean diameter of aerosol generator

Mean particle size (MPS) (3.0+/-0.3)μm, geometric std. deviation≤1.5

Andersen

I>7μm, II. 4.7-7μm, III. 3.3-4.7μm, IV. 2.1-3.3μm, V. 1.1-2.1μm

VI. 0.6-1.1μm

Size of aerosol chamber

600mm(Length)×80mm(Diameter)×5mm(Thickness)

Total particle number of positive control sampler

2200+/-500cfu

Ventilation flow of negative pressure cabinet

800m3/min

Operation Cabinet Size

740mm×940mmx440mm

Dimension of main machine

1200mm×650mm×2100mm (L×W×H)

     Power supply

AC220V+/-10%,50Hz

Noise

<65dB(A)

Weight

Approx. 150kg

Power

<1500W


Relative Products of XHF-107 Bacterial Filtration Efficiency (BFE) Tester

Hot lines:

+86-531-67880131

Address:No. 2 East Street Shengchan Road Jinan City Shandong China

Jinan Xinghua Instruments Co.,Ltd. All Rights Reserved