Your Location:Home > PRODUCTS>Environmental Testing Equipments

XHL-02QT Ampoule Break Point Tester

Product Description of XHL-02QT Ampoule Break Point Tester

Ampoule breakpoint tester is designed and manufactured in compliance to the DIN EN ISO <9187-1/2:2010> standard. According to this standard the break point of ampoules has to be tested to control the quality during production of empty ampoules and prior to the filling process.

Test Standards of XHL-02QT Ampoule Break Point Tester

Product Features of XHL-02QT Ampoule Break Point Tester

Industrial-grade 8-inch touch screen with a menu-based interface and flat design, eliminating unnecessary operations to access any interface with a single click.

A dedicated single-chip microcontroller measurement and control system for fracture force, featuring simple and fast operation, stable and reliable performance, and high precision.

The measurement and control software complies with the pharmaceutical GMP requirements for permission account management, facilitating standardized equipment usage by laboratory administrators;

The test curve displays the trend of experimental data changes throughout the entire testing process in real time;

High-precision motors work in conjunction with lead screw and rail structures to achieve vertical movement, ensuring lifting accuracy and enhancing testing precision.


Configure a high-precision force acquisition sensor capable of accurately measuring the maximum force at the moment of fracture.

The instrument features functions such as test result judgment, storage, querying, and data traceability tracking

The instrument features functions such as power-off memory, overload alarm, and shutdown protection to ensure test safety;

Equipped with a USB universal data interface for convenient data output and transfer;

Configure the mini printer to directly output and print test results;

Multi-level user permission management with password login for convenient user administration.


Technical Parameter of XHL-02QT Ampoule Break Point Tester

Sample

1-25ml

Force Testing Range

0-200N

Force Resolution

0.001N

Testing Accuracy

±0.5% of Indication Value

Testing Speed

 10mm/min (0.1-500mm/min adjustable)

Testing Speed Accuracy

 ±0.5% of Indication Value

Displacement Accuracy

 ±0.5% of Indication Value

Displacement Resolution

0.001mm

Test Span

36, 60mm

Control

Automatic, 8 Inch Touch Screen

Data Output

Printer, USB

Dimension

400(L)×300(B)×450(H)mm

Weight

18Kgs

Power supply

AC220V±10% 50Hz  Single Phase

Standard Configuration

Main Frame, Micro Printer, Test Fixture, Protection Cover, Broken Sample Collection Container)(Broken samples will slide into container), Print Papers

Optional Configuration

PC Software and PC 


Relative Products of XHL-02QT Ampoule Break Point Tester

Hot lines:

+86-531-67880131

Address:No. 2 East Street Shengchan Road Jinan City Shandong China