XHS-GT-V1/V3 Gas Permeability Tester is based on the differential pressure method, and is professionally applicable to the determination of gas transmission rate as well as solubility coefficient, diffusion coefficient and permeability coefficient of plastic films, composite films, high barrier materials, sheeting and aluminum foils.
ISO 15105-1, ISO 2556, GB/T 1038-2000, ASTM D1434, JIS K7126-1, YBB 00082003
Prepare baseline correction function, measure system error, ensure test accuracy;
Standard film correction function, sensor external calibration function;
Real-time monitoring of the test state, the whole process of saving the test state, the realization of all the test data curve can be reproduced, easy to evaluate the test process;
The test curves of each chamber can be displayed individually or in groups, which is convenient for comparison test. Test curve with full screen display, local observation, variable ratio display and other capabilities.
Power failure data saving function, data sampling rate is adjustable, automatic repeated test function;
Open database system for data analysis
The gas permeability, solubility coefficient and diffusion coefficient of the sample can be measured simultaneously
The three test chambers are completely independent and can test three same or different samples at the same time
Wide range, high precision temperature and humidity control, meet the test under various test conditions
It provides both proportional mode and standard mode test process judgment modes
The test range can be expanded according to needs to meet the requirements of high transmittance test
Data fitting can be carried out at any temperature, and the test results under extreme conditions can be easily obtained
Support toxic gas and flammable and explosive gas testing (customization needed)
The system is controlled by computer and the whole test process is completed automatically
Provide standard film for rapid calibration to ensure the accuracy and versatility of test data
Equipped with RS232 universal data interface (V3), convenient data transmission.
Model | XHS-GT-V1 | XHS-GT-V3 |
Control Mode | 10 Inch Touch Screen | PC |
Test Range | 0.01~50,000 cm3/m2·24h·0.1MPa (standard volume) | |
Number of Specimens | 1 | 3 (with independent test results) |
Main Sensor Quantity | 1 | 3 |
Vacuum Resolution | 0.01 Pa | |
Vacuum Degree of Test Chamber | <20 Pa | |
Temperature | 5°C ~ 65°C (5C -95C is optional) | |
Temperature Control Accuracy | ±0.05C (standard) | |
Temperature Resolution | 0.01C | |
Humidity | 0%RH, 5%RH-95%RH | |
Accuracy | ±1%RH | |
Specimen Size | Φ97 mm | |
Test Area | 38.48 cm2 | |
Test Gas | O2, N2 and CO2 (outside of supply scope) | |
Test Pressure | -0.1 MPa - +0.15 MPa | |
Vacuum tube diameter | 1/4" | |
Calibration mode | Calibration of standard film and external calibration of vacuum sensor | |
Sample thickness | ≤3mm | |
Gas Supply Pressure | 0.4 MPa ~ 0.6 MPa | |
Port Size | Φ6 mm PU Tubing | |
Instrument Dimension | 765 mm (L) x 440 mm (W) x 470mm (H) | |
Weight | 60Kgs | 88Kgs |
Power Supply | 220VAC 50Hz / 120VAC 60Hz | |
Standard Configuration | Instrument, Constant Temperature Control Device, Professional Software (V3), Computer (V3), Round Sample Cutter, Vacuum Grease, and Vacuum Pump | |
Optional | Optional: sampling blade, vacuum grease, vacuum pump oil, fast quantitative filter paper, humidity generator | |
Note: The air source of the machine is ф 6mm polyurethane pipe; gas source and distilled water are provided by users |